Getting Started
Page 2-3
Initial Configuration of the Model
2010D
Before the Model
2010D
can perform calibrations, it must be
configured for the intended application. Following is the list
of tasks that must be accomplished in order to place the unit in
proper working order.
[1] Set the Model
2010D
's time and date to local time as described
in the Subsection "Setting the Time and Date" in Section 6.
[2] Select the source gas and diluent gas. This is covered in the
"Installation" section of the manual which begins on page 4-1.
[3] Determine source and diluent port usage on the rear panel of
the Model
2010D
. This is covered in the "Installation"
section of the manual which begins on page 4-1.
[4] Set up the
2010D
for your specific calibrations. The "Model
2010D
Calibration Set-up" section, starting on page 6-1, covers this
topic.
[5] Verify and/or calibrate the flow controllers and the ozone
generator. This topic, covered in the section titled
"Calibrating Model
2010D
Components", starts on page 8-1.
System Operation and Maintenance Precautions
Care should be taken to leave factory set parameters unchanged.
Altering factory parameters can result in improper operation or even
damage to the unit. Improper use of the diagnostic routines can
also result in damage to the Model
2010D
.
When performing maintenance on the Model
2010D
, Power should be
removed. Tools and small loose parts could cause damage to the unit
if they come in contact with electrical circuits.
Anti static procedures should be used at all times when working with
electronic components.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......