Internal UV Ozone Photometer
Page 9-5
The voltage range of each output is independently selectable by
installing jumpers labeled JP1 through JP8 on the detector preamp
board, near the rear panel (see Figure 9-2). All outputs are set
to the 10V range by default, with all jumpers installed (oriented
vertically). Other ranges can be selected by removing jumpers,
as shown in the table, below:
Output 1
JP1
JP2
Output 2
JP3
JP4
Output 3
JP5
JP6
Output 4
JP7
JP8
Range
---------------------------------
In
In
0-10V
In
Out
0-5V
Out
Out
0-1V
Operation
There are three basic modes of operation for the Model 2010
D
's
UV Ozone Photometer. These are:
Audit Mode
In Audit Mode, the Photometer is used to
measure the ozone output of the Model
2010D
and serves as an independent audit
device.
Ozone Servo Mode
In
this
mode,
the
Photometer's
measurement is used to automatically
adjust
the
2010
D
's
internal
ozone
generator. This results in a precise
ozone concentration that is independent
of temperature, pressure, humidity and
other factors that can affect ozone
production.
Continuous Mode
Ordinarily, the Photometer is activated
only during ozone calibrations. In
Continuous Mode, however, the Photometer
will remain active continuously while
power is applied to the
2010D
This
feature allows the 2010
D
's Photometer to
be operated independent of the 2010
D
's
other functions as a stand-alone ozone
monitor.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......