Internal UV Ozone Photometer
Page 9-18
Derivative Term Constant
The Derivative constant is multiplied by
the integral term of the PID equation.
The derivative term represents the
current rate-of-change of the error.
The derivative term is normally small or
zero.
PID Filter
The PID Filter smoothes the Photometer's
measurements prior to the PID equation.
This
is
necessary
since
excessive
Photometer noise can cause the PID loop
to attempt to compensate for variations
that are merely cause by noise. The
value entered for the PID Filter is the
number of photometer cycles to use as
the filter's time constant.
Filter Holdoff
This is the number of photometer cycles
to wait after a change in the ozone
setpoint
before
beginning
the
PID
filter.
PID Holdoff
This is the number of photometer cycles
to wait after a change in the ozone
setpoint before activating the PID servo
control. This is normally set such that
the Photometer's natural response to an
ozone change has stabilized before the
PID
loop
is
activated
for
final
adjustments.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......