Operation
Page 5-3
Predefined Operator Stepped Manual Calibration
To perform a manual calibration using predefined or programmed
sequences, perform the following steps:
[1] From the Main Menu screen, select Sequences and press the
Enter button. A pop-up menu will appear.
[2] Select Run, and press the Enter button. Another pop-up
menu will appear that allows the choice of which type of
manual calibration you wish to use.
[3] Select Operator Stepped from the menu and press the Enter
button. A list of sequences that have been predefined or
programmed will appear (see Figure 5-1).
Figure 5-1 Manual Sequence Selection Menu
[4] Select the sequence containing the point you want to run.
The "Select a Point to Run" screen shown in Figure 5-2 will
appear, displaying the Sequence Name and Type, the Primary
Gas Name and concentrations for each point in the sequence
and a box entitled "Run Point Number", which will allow the
user to select a calibration point to run.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......