Introduction
Page 1-2
_________________________________________________________________
A complete calibration system.
The Model
2010D
Calibration System
makes use of embedded microprocessor technology to enhance the
accuracy and control features of the calibration system. The
Model
2010D
is designed to be used manually, automatically or
semi-automatically by remote control to conduct calibrations.
The unit features an optional internal ozone generator. An
optional permeation oven and UV absorption photometer are also
available.
Versatile programming options.
The calibrator has the capability to
produce and store in memory 20 calibration sequences, with up to 20
levels of source/dilution in each sequence. Users employ the front
panel membrane keypad or serial port to input calibration parameters.
A bright, easy-to-read, 25-line x 80-character display shows
calibration and system configuration information.
Additional features include continually updated display of dilution
and source information, local and remote calibration control,
eight-port output manifold with port connections for up to seven gas
analyzers, and portable and 19-inch rack mounting options.
The Model
2010D
's precision ozone generator is suitable for use
as a transfer standard for ozone calibrations or can be used for
performing nitrogen dioxide calibrations by means of gas phase
titration (GPT). The ozone generator eliminates the need for an
additional ozone transfer standard to conduct audits or perform
calibrations.
The Model
2010D
's ozone generator's advanced electronic
circuitry and propriety photo detector enable the ozone
generator output to be so stable and accurate that the need for
a photometer is eliminated.
The internal UV absorption photometer option is a complete UV ozone
analyzer that can be used independently for monitoring ozone
production.
The Model
2010D
external permeation oven option is an effective
tool for generating precise gas concentrations for calibrating a
wide range of gas analyzers. The large oven capacity
accommodates many types and sizes of permeation devices. An
internal clean air supply, precise mass flow controller and
temperature
controller
ensures
a
highly
accurate
gas
concentration process.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......