Model
2010D
Calibration Setup
Page 6-3
[2] Select Gas Table and press the Enter button. Select Delete
and press the Enter button. The pop-up screen Delete Gas
Table Entry will appear. By pressing Enter, a drop-down
list of the gases in the Gas Table will appear.
[3] Select the gas you want to delete and press the Enter
button. A screen with the prompt to delete or ignore will
appear.
If you wish to delete the gas, select yes and
press the Enter button. To ignore, select the No and press
the Enter button. This step may be repeated as necessary
to delete additional gases.
Set up Gas Standards
Each compressed gas cylinder attached to a source inlet port
must be set up as a "Gas Standard". Gas Standards may be set up
as follows:
[1] Select Gases from the main menu and press the Enter button.
A pop-up menu will appear.
[2] Select Gas Standards and press the Enter button. Another
pop-up menu will appear. This menu lists the following
options: View to view the Gas Standards, New to add a Gas
Standard, Edit to modify a Gas Standard and Delete to
remove a Gas Standard.
[3] To add a Gas Standard, select New and press the Enter
button. The New Gas Standard screen shown in Figure 6-2
will appear.
Figure 6-2 New Gas Standard Screen
[4] Enter a unique name for the Gas Standard, its serial
number, its expiration date and the carrier gas.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......