Model
2010D
Calibration Setup
Page 6-26
The Digital I/O Groups may be set up by the following procedure:
[1] Select Sequences under the main menu and press the Enter
button. A pop-up menu will appear.
[2] Select Remote Activation from the menu. A pop-up menu will
appear.
[3] Select Digital I/O Groups from the pop-up menu and press
the Enter button. The Edit Digital I/O Group Assignments
screen will appear, as shown in Figure 6-9.
Figure 6-9 The Edit Digital I/O Group Assignments Screen
[4] If a separate Abort Control Input is needed (that will be
used to terminate a sequence), initialize it by setting the
desired bit to "1". When finished, press Enter to proceed
to the next field.
[5] Identify the bits that will be used for Remote Sequence
Activation by setting the bits to "1", then press Enter to
proceed to the next field.
[6] Identify the bits that will be used for Remote Point
Activation by setting the bits to "1", then press Enter to
proceed to the next field.
[7] Identify the bits that will be used as Sequence Status
Outputs by setting the bits to "1", then press Enter to
proceed to the next field.
[8] Identify the bits that will be used as Point Status Outputs
by setting the bits to "1", then press Enter to proceed to
the next field.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......