Internal UV Ozone Photometer
Page 9-11
Sample Temperature (°C)
This is the temperature of the sample
gas in the measurement tube as measured
by the Photometer's temperature sensor.
Lamp Temperature (°C)
This is the temperature of the heated
Photometer lamp block.
Setup Options
The Model 2010
D
's photometer is adjusted, calibrated and all
parameters pre-set for optimal operation in the standard
configuration prior to delivery to the customer. As such it is
normally not necessary or desirable to modify the operating
parameters of the Photometer. There are circumstances, however,
when performance could be improved by adjusting one or more of
these parameters.
---
CAUTION
---
Making
changes
to
parameters
in
the
Photometer Parameters screens can cause the
photometer and/or ozone generator to operate
incorrectly or erratically and should only be
attempted with a thorough understanding of
the Photometer and the relationship of each
parameter. In any case be sure to record the
original factory settings before attempting
changes.
The photometer Parameters may be viewed or changed by selecting
"Devices" from the main menu, then "Ozone Photometer" and then
"Photometer Parameters". The Photometer Parameters are presented
on three pages. The first page is shown in Figure 9-4.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......