Installation
Page 4-15
The output module which activates the air source is configured
at the factory to provide switched 12 volts DC. When using an
alternate compressed air source, current draw should be limited
to 200 ma. Inductive loads such as solenoids and relays must
have snubber diodes in parallel with their coils. The diode
anode should be connected to J8-6.
The Air Source Control connector J8 on the rear panel of the
Model
2010D
is a female type DB-9 connector.
Table 4-10
Air Source Activation Port Connections
SIGNAL SIGNATURE
2010D
J8
__________
_________________
J8-1
CHASSIS GROUND
J8-2
AIR ON HIGH (Switched DC)
J8-6
AIR ON LOW (Analog Ground)
Photometer Analog Output Connections
If the Photometer option is installed and the Photometer's
output is to be monitored by a strip chart recorder, datalogger
or other external device, a cable must be made with the
connections listed in the Table, below.
Table 4-11 Photometer Analog Output Connections
2010D
J9
SIGNAL SIGNATURE
_______
_________________
J9-1
CHASSIS GROUND
J9-2
DAS OUT 1
J9-3
DAS OUT 2
J9-4
DAS OUT 3
J9-5
RESERVED
J9-6
CHASSIS GROUND
J9-7
GROUND
J9-8
GROUND
J9-9
GROUND
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......