Serial Communications
Page 7-30
_________________________________________________________________
Modbus Coil Registers, Read/Write
Register
Address
Signal Description
0-99
Start Sequence or indicate activity
100
Activate a Purge or indicate purge active
101
Stop all cals or indicate inactivity
200-224 Set or read digital IOs
Modbus Holding Registers of Floating Point Values
(Even Addresses: High word, Odd Addresses: Low Word)
Register
Address
Signal Description
Units
Read/Write
0
Source 1 Flow Monitor
SLPM
Read-Only
2
Diluent Flow Monitor
SLPM
Read-Only
4
Photometer Ozone Measurement
PPB
Read-Only
6
Ozone Generator Monitor
mV
Read-Only
8
Ozone Generator Flow
SLPM
Read-Only
10
Ozone Generator Lamp Setpoint
mV
Read-Only
12
Ozone Generator Temperature
°C
Read-Only
22
Perm Oven Temperature
°C
Read-Only
24
Perm Oven Flow
SLPM
Read-Only
26
Photometer Detector, Sample
mV
Read-Only
28
Photometer Detector, Reference
mV
Read-Only
30
Photometer Sample Flow
SLPM
Read-Only
32
Photometer Lamp Temperature
°C
Read-Only
34
Photometer Sample Pressure
InHg
Read-Only
36
Photometer Sample Temperature
°C
Read-Only
38
Photometer Cal Factor Slope
NA
Read/Write
40
Photometer Cal Factor Intercept
PPB
Read/Write
50
Source 2 Flow Monitor
SLPM
Read-Only
52
Source 1 Flow Control
SLPM
Read/Write
54
Source 2 Flow Control
SLPM
Read/Write
56
Diluent Flow Control
SLPM
Read/Write
58
Ozone Generator Control
PPB
Read/Write
60
Ozone Generator Set Voltage
mV
Read/Write
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......