Serial Communications
Page 7-9
_________________________________________________________________
Response Start
The character that initiates a response
when data is returned. The <CR> character
is the default for the ML protocol.
Acknowledge
The response character that acknowledges
the execution of a command when no data is
returned. The <ACK> character is the
default for the ML protocol.
Negative Acknowledge
The response character that indicates that
there was a problem responding to the
command. The <NAK> character is the
default for the ML protocol.
Command End
The end-of-command character.
The <CR>
character is the default for the ML
protocol.
Field separator
The character used to separate parameter
or data fields in commands and responses.
The comma character is the default for the
ML protocol.
Configuring the Computer for Terminal Mode
The computer's terminal emulation program must be configured
before a successful remote connection can be made. Following
are several considerations and descriptions of how to set up
each item in Procomm for DOS. For more details, refer to the
terminal program's documentation.
The following items can be configured in Procomm for DOS by
entering the Line/Port Setup screen with the ALT-P command:
Comm Port
Select the comm port (usually COM1 or
COM2) to which the Model
2010D
or modem is
attached.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......