Introduction
Page 1-5
_________________________________________________________________
Ozone Generator:
Output range 0.05 - 1.0 PPM at 5 SLPM instrument flow (Optional
extended output range 0.05 - 1 PPM at 10 SLPM instrument flow)
Ac 2% of set point or + 3 PPB at 5 SLPM
Nominal ozone flow of 100 SCCM maintained 1 SCCM by
precision flow controller
UV lamp maintained at 50 + 0.1 oC
Optional UV optical servo control loop
Optional UV absorption photometer
Internal UV Absorption Photometer Option
:
Ozone monitoring ranges 100 PPB to 20 PPM full scale
Lin 1 PPB or + 1 percent of full scale, whichever is
greater
Pre 1 PPB
Lower detectable limit 0.8 PPB
Zero drift for 24 hours and 30 days less than 1 PPB
Span drift for 24 hours and 30 days less than 0.5 percent of reading
Lag time less than 10 seconds
Rise and fall time of 95 percent of final reading less than 60 sec
at 500 SCCM sample flow rate
Nominal flow rate of 500 SCCM monitored by mass flow meter
Internal sample vacuum pump
Recorder and data acquisition system analog outputs
Operating temperature 5o - 40 oC
Calibration Definitions:
Twenty user defined calibration sequences with up to 20 points per
sequence
Calibrations may be gas blends, ozone or GPT
Two diluent gas definitions and 20 source gas standard definitions
Twenty timer-driven calibration routines that perform predefined
calibration sequences on a calendar of events
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......