7
Display
result
FZ3 Processing Items List Manual
Display Last NG Image
429
Display Last NG Image
Data, images and drawn data (up to 4 sets) for a NG based on NG conditions defined using an
expression can be saved.As the saved image is stored in memory, it is maintained even if operations are
performed in the window.
Used in the Following Case
●
Image and data for NG case are saved.
NG Error Judgement (Display Last NG Image)
Sets conditions for NG judgement.
1.
Tap [NG judgement] in the Item Tab area.
2.
Set the judgement mode in the "NG settings" area.
Setting item
Set value
[factory
default]
Description
Judgement
mode
[One NG]
The image is saved even if only one of the judgement criteria set
using "Judgement expression" has a judgement of NG.
All NG
The image is saved if all of the judgement criteria set using
"Judgement expression" have a judgement of NG.
Acesst 1.0 6251194