Measurement
64
Sensitive Search
FZ3 Processing Items List Manual
Setting item
Set value
[factory default]
Description
Keep disabled
setting
●
[Checked]
●
Unchecked
When the model is registered, this holds the disable setting for
the sub-region set during the last time the model was
registered.
Save
registered
model.
●
Checked
●
[Unchecked]
To save the entire image used for model registration, place a
check at this option.
4.
Tap [OK].
The model is registered.
Changing Model Parameters
Model parameter values can be changed as needed to address unstable measurement results or to
increase the processing speed.Normally, the factory default value will be used.
After changing a setting, re-register the model.
1.
In the "Model parameter" area, set each item.
Setting item
Set value
[factory default]
Description
Rotation
●
Checked
●
[Unchecked]
When the measurement object is rotating, place a
check at "Rotation" and specify how many degrees
the model created rotates each time and through
what range of angles.A smaller skipping angle
increases stability, but slows down the
processing.The normal direction is clockwise.
Angle range
[-180 to 180]
Skipping
angle
1 to 30
[5]
Smart mode
●
[Checked]
●
Unchecked
Checking the "Smart mode" option enables a
high-speed rotation search.However, the stability
may be lowered when the model shape aspect
ratio is large or when the NOT mask is used.
Stability
1 to 15
[The default
value depend on
the connected
camera.9 or 12]
Specify which is to have priority, measurement
stability or speed.
If lowering stability does not speed up processing,
it is likely that many candidates have been
detected.In this case, specify a larger value for
"Candidate level" or "Stab."
Preciseness
1 to 3
[2]
Specify which is to have priority, measurement
positional precision or speed.
Changing Sub-model parameter
Set the "Sensitive model setting" as necessary.
Acesst 1.0 6251194
Summary of Contents for XPECTIA FZ3
Page 11: ...FZ3 Processing Items List Manual 9 Index 435 ...
Page 39: ...1 Input image FZ3 Processing Items List Manual Camera Image Input HDR 37 ...
Page 41: ...1 Input image FZ3 Processing Items List Manual Camera Switching 39 ...
Page 398: ...4 Support measurement 396 Iris FZ3 Processing Items List Manual ...
Page 410: ...5 Branch 408 DI Branch FZ3 Processing Items List Manual ...