![background image](http://html.mh-extra.com/html/omron/xpectia-fz3/xpectia-fz3_user-manual_1633321416.webp)
6
Output
result
414
Data Output
FZ3 Processing Items List Manual
Important
About output when Ethernet is set as the output destination
●
Output format: ASCII
1 packet is output for each 1 unit of serial data output.
When multiple units of serial data are output, that many packets are output.
●
Output format: binary
1 packet is output for each 1 data item of serial data output.
●
Set an appropriate offset value according to the maximum value of write address for the PLC. If a value
exceeding the maximum value permitted by the PLC is set, an error occurs. (FZ3-9
/H9
only)
Key Points for Test Measurement and Adjustment (Data Output)
The following content can be confirmed in the "Detail result"area using text.
Displayed items
Description
Judge
Judgement result
Expression0-7
Results of expressions 0 - 7
Measurement Results for Which Output Is Possible (Data Output)
The following values can be output using processing items related to results output. It is also possible to
reference measurement values from expressions and other processing units.
Measurement item
Character string
Description
Judgement
JG
Judgement result
Data 0 to 7
D00 to D07
Results of expressions set for output data 0 to 7
Acesst 1.0 6251194
Summary of Contents for XPECTIA FZ3
Page 11: ...FZ3 Processing Items List Manual 9 Index 435 ...
Page 39: ...1 Input image FZ3 Processing Items List Manual Camera Image Input HDR 37 ...
Page 41: ...1 Input image FZ3 Processing Items List Manual Camera Switching 39 ...
Page 398: ...4 Support measurement 396 Iris FZ3 Processing Items List Manual ...
Page 410: ...5 Branch 408 DI Branch FZ3 Processing Items List Manual ...