398
10 .2 Error codes and countermeasures
Error code
Displayed message
Problems/countermeasures
S008**
S00900
S00910
S00920
S00930
[AeroDR Stitching Unit name] MCB Error
Contact a service representative .
Contact Konica Minolta technical representatives .
S00940
S00A00
S00A10
S00A20
[AeroDR Stitching Unit name] Motor Error
Please power OFF/ON of AeroDR Stitching unit .
Restart the AeroDR Stitching System .
S00A30
[AeroDR Stitching Unit name] Sensor Error
Please power OFF/ON of AeroDR Stitching unit .
S00A40
S00A50
[AeroDR Stitching Unit name] Motor Error
Please power OFF/ON of AeroDR Stitching unit .
S00A60
S00C90
[AeroDR Stitching Unit name] Sensor Error
Please power OFF/ON of AeroDR Stitching unit .
S00CA0
[AeroDR Stitching Unit name] Side Cover Open
Please close the AeroDR stitching unit cover and
press RESET switch .
Close the side cover of the AeroDR Stitching Unit, and
press the Reset switch on the AeroDR Stitching X-Ray
Auto-barrier Unit to cancel the error .
S00CB0
[AeroDR Stitching Unit name] Panel Error
Properly set the grid inside AeroDR stitching unit,
close the over and press the RESET switch .
Check that the DR Detector is correctly inserted in the
AeroDR Stitching Unit, and close the side cover . Press
the Reset switch on the AeroDR Stitching X-Ray Auto-
barrier Unit to cancel the error .
S00CC0
[AeroDR Stitching Unit name] Grid Error
Properly set the grid inside AeroDR stitching unit,
close the over and press the RESET switch .
Check that the grid is correctly installed on the AeroDR
Stitching Unit, and then close the side cover . Press
the Reset switch on the AeroDR Stitching X-Ray Auto-
barrier Unit to cancel the error .
S00CD0
[AeroDR Stitching Unit name] Panel Error
Properly close the inner cover of AeroDR stitching unit
and switch the power OFF/ON .
(1) Open the side cover on the AeroDR Stitching Unit
and close the internal lid .
(2) Close the side cover and restart the AeroDR Stitch-
ing System .
S200**
[AeroDR Stitching Unit name] Internal connection error
Please power OFF of AeroDR Stitching unit .
Please confirm the connectivity of Power box and auto
barrier unit before switching the power to ON .
(1) Turn off the AeroDR Stitching System.
(2) Check the power unit and the AeroDR Stitching X-
Ray Auto-barrier Unit are connected by the power
cable, and turn on the AeroDR Stitching System .
S208**
[AeroDR Stitching Unit name] Internal Error
Please power OFF/ON of AeroDR Stitching unit .
Please confirm the connectivity of Power box and
Stitiching unit before switching the power to ON .
(1) Turn off the AeroDR Stitching System.
(2) Check the power unit and the AeroDR Stitching
Unit are connected by the power cable, and turn
on the AeroDR Stitching System .
S20C50
[AeroDR Stitching Unit name] Network Error
Please confirm the network of AeroDR stitching unit
and CS-7, switch on the power OFF/ON of AeroDR
stitching unit .
Check that the AeroDR Stitching Unit and this device
are connected by the network cable and restart the
AeroDR Stitching System .
S20C60
[AeroDR Stitching Unit name] Network Error
Please confirm the network of AeroDR stitching unit
and CS-7, press the AeroDR Stitching unit RESET
switch .
Check that the AeroDR Stitching Unit and this device
are connected by the network cable, and press the
Reset switch on the AeroDR Stitching X-Ray Auto-
barrier Unit to cancel the error .
S30460
[AeroDR Stitching Unit name] Height detection (out of
range)
Please press the AeroDR Stitching unit RESET switch .
Repeat the height detection .
(1) Press the Reset switch on the AeroDR Stitching
X-Ray Auto-barrier Unit to cancel the error .
(2) Check the collimator installation position in the X-
ray device is correct .
(3) Execute height detection.
(4) If the problem is not solved by the above method,
there is a possibility that the intensity of the expo
-
sure field lamp has changed. Change the bright
-
ness setting .
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...