388
10 .2 Error codes and countermeasures
Error code
Displayed message
Problems/countermeasures
C36022
At PrimtComposer process, processing is timeout .
(1) Try performing the operation again .
(2) Contact Konica Minolta technical representatives .
C37001
[DR Detector name] Connecting detector
Check the network equipment and communication
cable connection status .
C37002
[DR Detector name] disconnection error
C37003
C37004
Error . Please touch the work area .
Check the network equipment and communication
cable connection status .
Touching either the working area or exposure condi
-
tion items may fix the problem.
C37005
[DR Detector name] Disconnected energy saving not
set . Please touch the work area .
(1) Touch either the working area or exposure condi
-
tion items .
(2) Check the network equipment and communication
cable connection status .
(3) Stop the examination.
C37006
[DR Detector name] wireless forbidden error
Check the network equipment and communication
cable connection status .
C37007
[DR Detector name] version up error
Version upgrade failed .
Check the connection and restart the AeroDR cas-
sette .
C37008
[DR Detector name] calibration error
Calibration failed .
Check the connection and retry .
(1) Check the network equipment and communication
cable connection status .
(2) Restart the calibration tool .
C37009
There are images unreceived to the AeroDR .
Do not turn off the power of AeroDR.
Return AeroDR detector to previous connection
state(Wired/Wireless) . In the case of wireless connec-
tion, the electric wave state has deteriorated .
Please check an electric wave state by the AeroDR
icon at the upper right of a screen, and wait until con-
nection returns .
(1) Check the network equipment and communication
cable connection status .
(2) Press [OK], and wait until image re-acquisition is
complete .
C37010
[DR Detector name] Correction processing error
Correction processing error .
Restart the CS-7 and retake .
Restart this device and perform the exposure again.
C37012
[DR Detector name] Additional registration error
Check the network equipment and communication
cable connection status .
C37013
[DR Detector name] Registration cancelation error
C37014
[DR Detector name] Set panel data error
C37015
[DR Detector name] Get panel data error
C37017
[DR Detector name] Attribute information error.
C37018
[DR Detector name] Error in image file creation.
Please contact the service representation .
Contact Konica Minolta technical representatives .
C37019
[DR Detector name] Exposure Permission Setting Er
-
ror
Please touch the work area .
(1) Touch either the working area or exposure condi
-
tion items .
(2) Check the network equipment and communication
cable connection status .
C37025
[DR Detector name] Exposure Mode Setting Error
Please touch the work area .
Touching either the working area or exposure condi
-
tion items may fix the problem. If not, contact Konica
Minolta technical representatives .
C37054
Exposure field setting error.
Check the network connection and touch this expo
-
sure condition .
(1) Check the network equipment and communication
cable connection status .
(2) When [Ready] is displayed, press the Exam. Tag.
C37055
Communication error .
Check the network connection and touch this expo
-
sure condition .
C37056
Version up failed .
Check the connection and restart the AeroDR XG BOX/
AeroDR Interface Unit2 .
(1) Check the network equipment and communication
cable connection status .
(2) Restart the device and the AeroDR Interface Unit .
C37058
Failed to set the Setting data .
Please check network connection and retry .
Contact Konica Minolta technical representatives .
C37059
Failed to get the Setting data .
Please check network connection and retry .
Contact Konica Minolta technical representatives .
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...