8.2 Dialog boxes/balloons and countermeasures
Displayed message
Problems/countermeasures
Registration Failed
This is displayed when registering of patient information has
failed .
Re-reception Error . Check network condition and Retry opera-
tion .
Displayed when image reacquisition fails .
After a few seconds, press [Resend] for re-acquisition.
Re-reception Error . Check network condition and touch applica-
tion bar after few seconds .
Displayed when image reacquisition fails .
Check the connection status of the network device and commu-
nication cables, and tap the Application Bar after a few seconds
or tap the Exam. Tags (in the yellow frame) to re-acquire the
image .
Re-reception Error . Check network condition and touch applica-
tion bar after few seconds .
Image re-reception from the AeroDR Stitching System failed .
After a few seconds, touch the yellow frame on the Application
Bar, and receive the image again .
Restoring Previous Setting . Please Wait…
Displayed when area switching fails and the area settings are
returned to their original status .
Same Patient ID: [0],Patient Name: [1],Sex: [2],DOB:[3] avail
-
able in Database, do you want to overwrite?
This is displayed when registering patient information that is
already registered in this device .
Scheduled Date(From) cannot be greater than Scheduled
Date(End) .
This is displayed when the value for the examination schedule
date entered in (From) is larger than that for (To) .
Enter appropriate values .
Search Error occurred
This is displayed when a search to the higher system has failed .
Select Folder Path .
Displayed when a folder path is not selected .
Select a folder path .
Send Cancel Error occured . Please refresh the study list and try
again .
Displayed when attempting to cancel transmission of a study
which does not exist.
Confirm that the target study has not been moved to another
device .
Send Error occured . Please refresh the study list and try again . Displayed when attempting to transmit a study which does not
exist or which has already been reserved for transmission to
another device .
Confirm that the target study has not been moved to another
device .
Separately, if you want to use the panel Aero-Sync environ-
ment was conducted calibration must be calibrated Aero-Sync
environment .
This is displayed when using the DR Detector calibrated in Sync
Mode in the Aero-Sync environment .
Perform calibration in the Aero-Sync environment .
Separately, if you want to use the panel Sync environment was
conducted calibration must be calibrated Sync environment .
This is displayed when using the DR Detector calibrated in
Aero-Sync Mode in the Sync environment .
Perform calibration in the Sync environment .
Setting File loading failed due to file lock
The file may be corrupted.
Load the file again.
Sex is mandatory, Please enter Sex.
Displayed when mandatory fields have not been filled in.
Press [OK] to close the screen.
Some of the components are going to expire or have expired.
Please Check .
Displayed when there is a unit which will require or does require
an exchange. Contact Konica Minolta technical representatives.
Space Not Enough . Please Check Media
This is displayed when there is not enough space to write to the
media .
Start Date should be less than End Date .
Displayed when the end date is set earlier than the start date .
Set the start date earlier than the end date .
Start value should be less than Stop value .
Displayed when setting the Start value to a number higher than
the Stop value .
Enter a Start value that is lower than the Stop value .
Stitched image will be selected . Is it OK to clear selected
image(s) .
Displayed when trying to decompress a combined image into
the stitch adjustment panel. Press [OK] to decompress it.
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...