5.3 Exposure with DR Detector (Aero Sync mode)
2
After positioning the patient, check the X-
ray detection level, then press [AeroSync] .
• If the control panel does not switch to
[AeroSync], press the [AeroSync] tab.
• Change the X-ray detection level as needed .
• After checking the X-ray detection level, press
[AeroSync].
• A message "Ready" is displayed in the mes-
sage/button area . The screen is locked .
• The screen is locked . To return to the previous
screen to change the position of the patient,
etc., press [Cancel].
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• To prevent incorrect operations, always press
[AeroSync] and check that "Ready" is displayed.
• Conditions for changing the detection level vary
depending on the X-ray device being used . If a survey
has been carried out when the system was installed,
follow the survey results .
• When the AeroDR power management level is set to
[3], [4] or [5], it may take some time until the exposure
becomes ready after you have pressed [Aero Sync]. In
this case, [Standby] button instead of [Cancel] button
is displayed in the message/button area .
•
After pressing [AeroSync], do not touch the wired
connector of the DR Detector until the image transfer
process completes .
• When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
• The X-ray detection level can be set when making an
exposure. Refer to "5.3 Exposure with DR Detector
(Aero Sync mode)" for information on the X-ray detec-
tion levels that can be set .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
3
Perform exposure .
• Configure the X-ray device settings and check
the exposure setting before making an expo
-
sure .
Press the exposure switch
•
As soon as the exposed image is transferred,
the image is displayed in the image display area .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When the "Ready" is displayed, immediately perform
exposure.
• Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
•
If image acquisition started without performing ex
-
posure, this may result in incorrect detection due to
noise, etc . Perform rejecting (image deletion) and re-
peat exposure.
•
If image acquisition does not start after exposure, the
voltage may not be enough. Check the exposure con
-
ditions and repeat exposure.
•
When exposure is done while "Ready" is not dis
-
played, have an interval of 1 minute or more before
repeating exposure.
• When linked with the X-ray device, because the se-
lected exposure conditions at the time of X-ray expo
-
sure are linked to the irradiation results, always check
the exposure conditions before X-ray exposure.
• Depending on the type of X-ray device, several sec-
onds may elapse before notification of the irradiation
results . For this type of model, take care because
sometimes irradiation results are not reflected correct
-
ly if the exposure conditions are changed immediately
after X-ray irradiation .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...