401
10 .2 Error codes and countermeasures
Chapter 10
Error code
Displayed message
Problems/countermeasures
LAMPD
Erase lamp life expired.
The Erasure Lamp need to be replaced .
Contact Konica Minolta technical representatives .
LAMPG
Erase lamp life had exceeded. Please call service to
get lamp changed .
LASER1
Laser Error
Contact Konica Minolta technical representatives .
POLI1
Polygon PLL Error
SIG1
Uses Default Shading Coefficient
SIG2
Uses Default QRV Table
SIG3
Uses Default Shading Coefficient / QRV Table
0001
Error occurred while copying setting files
The setting file may have not been updated. Check
the saved settings, and perform the update again .
04001 - 05000
-
(1) Restart the REGIUS series .
(2) If the error persists after restarting, restart the
device and the REGIUS series .
05001
JM Error . Please call service . Or change to Backup
JM if it exists.
(1) When there is a backup JM, switch the JM .
(2) If there is no backup JM, contact Konica Minolta
technical representatives .
05002
JM Switching Error
Check that the JM in the backup destination is run-
ning, and switch the JM again .
05003
ExchangeParts DB
Initialization Error
Check that the JM is running, and restart the device .
05004
JM switched .
JM switching was completed successfully .
05005
Please restart Regius
JM switching was completed successfully .
Restart all CR cassette readers .
05006
JM Error . Please call service . Or change to Primary
JM
(1) Switch from the backup JM to the primary JM .
(2) Contact Konica Minolta technical representatives .
06000 -
06785
-
(1) Restart the REGIUS series .
(2) If the error persists after restarting, restart the
device and the REGIUS series .
07000
Initialization failed . Cassette is set in the reader .
Please open the door and remove the Cassette and
Plate .
(1) Remove the cassette and plate from the device .
(2) Delete the image on the removed plate using the
Erase mode .
07001
07002
Initialization failed . Plate still might be inside the
reader . Please open the door and check inside .
07003 - 07005
Initialization failed . Cassette is set in the reader .
Please open the door and remove the Cassette and
Plate .
07006
Initialization failed . Plate still might be inside the
reader . Please open the door and check inside .
07007 - 07016
Initialization failed . Sensor Error was detected during
initialization .
Open and close the top door . Perform initialization of
the device again .
Check if the error reoccurs .
07017
Error was detected while processing the cassette .
Please open the door and remove the Cassette and
Plate . After removal, erase the cassette in ERASE
MODE .
(1) Remove the cassette and plate from the device .
(2) Delete the image on the removed plate using the
Erase mode .
07018
07019
Initialization failed . Sensor Error was detected during
initialization .
Open and close the top door . Perform initialization of
the device again .
Check if the error reoccurs .
07020
Initialization failed . Cassette was detected . Please
open the top door and remove the Cassette .
Open the top door and remove the cassette .
07021
Initialization failed . Sensor Error was detected during
initialization .
Open and close the top door . Perform initialization of
the device again .
Check if the error reoccurs .
07022
Single unit operation error .
Contact Konica Minolta technical representatives .
07023
07024
Initialization failed . Sensor Error was detected during
initialization .
Continued on the next page
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...