389
10 .2 Error codes and countermeasures
Chapter 10
Error code
Displayed message
Problems/countermeasures
C37060
Exposure sync setting error.
Check the network connection and touch this expo
-
sure condition .
Check the network equipment and communication
cable connection status .
C37064
Network connection between AeroDR and AeroDR XG
BOX/AeroDR Interface Unit2 may not be good .
Check the network connection and take exposure
once again .
The DR Detector and AeroDR Interface Unit2, AeroDR
Generator Interface Unit or AeroDR Generator
Interface Unit2 communications status is poor .
Check the network status and then try performing the
exposure again.
C37102
5th AeroDR Detector can not be registered .
Up to four DR Detectors can be registered to one ex
-
posure room. A fifth DR Detector cannot be registered
because four panels are already registered .
C37103
Unknown AeroDR Detector .
The DR Detector is unregistered .
Registered DR Detectors are available .
C37104
Unknown exposure device.
Exposure device is not registered.
Registered panels (DR Detectors or CR cassettes) are
available .
C37111
An error occurred when canceling registration of the
AeroDR panel .
Check the network equipment and communication
cable connection status .
C37116
The license of the panel you tried to register does not exist
or the number of licenses is insufficient. Obtain a license or
register the panel for the number of licenses obtained .
(1) Acquire an DR Detector license .
(2) Register DR Detectors without exceeding the num
-
ber of licenses .
C37119
Non-activated AeroDR detector is found .
Contact Konica Minolta technical representatives .
C37121
There’s no selectable channel.
It will take maximum 30 minutes to recover the wire
-
less LAN .
This appears when there are no accessible chan-
nels available when attempting to switch access point
channels while detecting radar waves .
Please wait until wireless communication is restored .
C37122
Radar wave has been detected . Changing the com-
munication channel .
It takes about 1 minute for channel change .
This appears when radar waves are detected .
Please wait until the device has finished switching the
access point channel .
C37123
Channel has been changed .
This appears when the device has finished switching
the access point channel .
C37124
Failed to start the AP Process . Please restart the PC . Restart the device . If this does not solve the problem,
contact Konica Minolta technical representatives .
C37150
[LongLenIF] Initialization failed.
Contact a service representative .
(1) Restart the device .
(2) Contact Konica Minolta technical representatives .
C37151
[LongLenIF] Termination failed.
Contact a service representative .
C37153
[DR Detector name] Radiation field detection Failed.
Radiation field light got off.
Check the light intensity and exposure range, and
perform exposure field detection again.
C37154
[DR Detector name] Radiation field detection Failed.
Radiation field is out of range.
C37155 -
C37157
[DR Detector name] Radiation field detection Failed.
C37300
Termination status
(1) Please check the connection status .
(2) Restart the device .
C3730*
C3731*
C3732*
Initialization error
C3733*
C3734*
Conversion error
Continued on the next page
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...