Keysight B1505A Configuration and Connection Guide
7-35
Connection and Ordering Examples
Configuration Examples for Lateral Device Measurement with Wafer Prober
3 kV, 40 A Measurement for On-Wafer Lateral Device
•
Device Type: Lateral MOSFET (4 terminals)
•
Key Specification
•
Max Voltage/Current: 3 kV/30 A with Module Selector, 3 kV/40 A without Module
Selector
•
Module Selector for multi-parameter automated measurement function
Figure 7-18
Connection example for 3kV, 30A measurement with N1258A Module Selector
HCSMU
Dual HCSMU
Combina
Ɵ
on
adapter
16493S-021
HVSMU
HCSMU
GNDU
MCSMU
B1505A
N1258A
MPSMU
Protec
Ɵ
on
Adapter
N1261A-003
D
G
S
HCSMU Non-
Kelvin adapter
16493S-011
HF
LF
R-Box
1kohm
N1262A-010
For Gate
16493G-001 Digital I/O cable (Included in N1258A)
N1254A-104 TRX(J) to BNC(P) adapter
F
S
F
S
F
S
High F
High S
Low S
Low F
Digital I/O
HV SMU
HCSMU F
HCSMU S
GNDU
F
S
16493J Interlock cable (Included in B1505A)
For Chuck
Chuck
For Drain
For Drain
For Drain
16493T (Included in the HVSMU)
16493S (Included in the HCSMU)
16494A (Included in the MCSMU)
16493L (Included in B1505A)
16494A-003
16493S (Included in the HCSMU)
30 cm HCSMU cable
(Included in 16493S-021)
16493T
16493U-001
1250-2405
16494A
(Included in the MPSMU)
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......