Keysight B1505A Configuration and Connection Guide
5-33
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect HVSMU/HCSMU Fast Switch
Table 5-23
To Connect N1267A Input
Table 5-24
To Connect N1267A Output
CAUTION
The GNDU can sink current of up to 4.2 A. Use 16493L GNDU cable to connect the
GNDU to the N1267A.
Do
not
use normal triaxial cable (16494A) because the maximum current rating of the
cable is 1 A.
CAUTION
Never connect the HVSMU Force and Guard terminals to any output, including circuit
common, chassis ground, or any other measurement resource such as SMU. Connecting
other measurement resource may damage the connected one.
NOTE
Do not connect or put any conductor on the HCSMU Low Force and Low Sense terminals,
outer conductor of the coaxial connectors. Connecting or putting conductor of circuit
common, chassis ground, or any potential on causes the measurement error.
B1505A terminals
Required cables
N1267A terminals
HCSMU
Force
16493S HCSMU cable, 3 m or 1.5 m, 1 ea.
HCSMU
Force
Sense
Sense
HVSMU
Force
16493T HVSMU cable, 3 m or 1.5 m
HVSMU
GNDU
16493L GNDU cable, 3 m or 1.5 m
GNDU
MCSMU
Force
16494A Triaxial cable, 3 m or 1.5 m
MCSMU
Force
Sense
16494A Triaxial cable, 3 m or 1.5 m
Sense
N1267A
terminals
Required cables/Description
Connect to
Low
Force
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
DUT low terminal
(ex: source)
High
Force
Coaxial cable with HV(plug) connector. N1254A-506 can be used.
Use coaxial cable to minimize affect of inductance.
DUT high terminal
(ex: drain)
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......