1-54
Keysight B1505A Configuration and Connection Guide
Configuration Guide
Accessories for Wafer Prober and Your Own Test Fixture
16493S-021
Adapter, 40A, select this if 16493S-020 is
not used
•
This adapter is used with the N1259A
Test Fixture, N1258A Module Selector,
or 16493S-010/011 HCSMU Adapter
•
If a N1258A Module Selector is used
with this adapter, the maximum current
is limited to 30 A. If 40 A capability is
needed, please connect this adapter to
device directly.
•
Output connectors are compatible with
B1512A HCSMU.
•
The 16493S-021 is furnished with one
30cm Triaxial cable and one 30 cm
High current coaxial cable for output
connection.
•
Do not connect or put any conductor on
the HCSMU Low Force and Low Sense
terminals, outer conductor of the
coaxial connectors.
•
Connecting or putting conductor of
circuit common, chassis ground, or any
potential on causes the measurement
error.
For Ultra High Current Measurement
N1254A-524
Ultra High Current Prober System Cable
•
The N1254A-524 cable is furnished
with one transparent shield with a slit.
•
Maximum current range is about 800 A
or more.
•
Cable length is 1.8 m.
Model/Option Description
Additional Information