1-8
Keysight B1505A Configuration and Connection Guide
Configuration Guide
B1505A Modules and Mainframe Options
Table 1-5
Mainframe Options
Description
Model/
Option
Note
Mainframe
Power Device Analyzer/Curve Tracer
B1505A
•
See
for furnished
accessories
Select modules (optional)
High Power Source/Monitor Unit module
(HPSMU), 200 V/1 A
B1510A-FG
•
Furnished with two triaxial cables
(16494A). Cable length 1.5 m or 3.0 m is
specified by 015 or 030.
•
Up to four modules can be installed.
•
The total number of installed HPSMU and
HCSMU modules cannot exceed 4.
Medium Power Source/Monitor Unit module
(MPSMU
)
, 100 V/100 mA
B1511B-FG
•
Furnished with two triaxial cables
(16494A). Cable length 1.5 m or 3.0 m is
specified by 015 or 030.
•
Up to ten modules can be installed.
High Current Source/Monitor Unit module
(HCSMU), 20 A at 20 V
B1512A-FG
•
Furnished with one pair of high current
cables (16493S) and Kelvin adapter
(16493S-010). Cable length 1.5 m or 3.0 m
is specified by 015 or 030.
•
Up to two modules can be installed.
•
The total number of installed HPSMU and
HCSMU modules cannot exceed 4.
High Voltage Source/Monitor Unit module
(HVSMU), 3000 V at 4 mA
B1513C-FG
•
Furnished with one HVSMU cable
(16493T). Cable length 1.5 m or 3.0 m is
specified by option 015 or 030.
•
Up to five modules can be installed.
Medium Current Source Monitor Unit
module (MCSMU), 1 A at 30 V
B1514A-FG
•
Furnished with two triaxial cables
(16494A). Cable length 1.5 m or 3.0 m is
specified by 015 or 030.
•
Up to six modules can be installed.
Multi Frequency Capacitance Measurement
Unit module (MFCMU)
B1520A-FG
•
Furnished with one CMU cable (N1300A).
Cable length 1.5 m or 3.0 m is specified by
option 015 or 030.
•
Up to one modules can be installed.
Specify the cable length (mandatory)
1.5m cable
015
•
Same cable length is applied to all
furnished cables.
3.0m cable
030
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......