Keysight B1505A Configuration and Connection Guide
5-9
Connection Guide for Wafer Prober and Your Own Test Fixture
Connection Overview
Table 5-5
To Make UHC Measurement Path
B1505A terminals
Required cables
N1265A terminals
MCSMU
1
1. HCSMU is substitutable. Then use 16493S HCSMU cable, N1254A-103 adapter for Force, and
N1254A-517 adapter for Sense.
Force
16494A Triaxial cable, 1.5 m, 1 ea.
UHC
Input
I Control
Force
Sense
16494A Triaxial cable, 1.5 m, 1 ea.
Sense
MCSMU
1
Force
16494A Triaxial cable, 1.5 m, 1 ea.
UHC
Input
V Control
Force
Sense
16494A Triaxial cable, 1.5 m, 1 ea.
Sense
MCSMU
1
Force
16494A Triaxial cable, 1.5 m, 1 ea.
Gate
Force
Sense
16494A Triaxial cable, 1.5 m, 1 ea.
Sense
GNDU
16493L GNDU cable, 1.5 m, 1 ea. Also see
GNDU
Digital I/O
16493G Digital I/O cable, 1.5 m, 1 ea. Also see
.
Digital I/O Input
Terminals
Required cables/Description
Connect to
Low Force
Cable with banana connector. N1254A-523 can be used.
DUT low terminal
(ex: source)
2
2. Non-Kelvin connection is not allowed. GNDU Sense is internally connected to Low Sense.
Low Sense
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
High Force
Cable with banana connector. N1254A-523 can be used.
DUT high terminal
(ex: drain)
2
High Sense
Triaxial cable with HV(plug) connector. N1254A-505 can be used.
Gate
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
DUT gate or base
LINE
Power cable, 1 ea., for connecting to power line
site AC power
Earth
Wire, 1 ea., for earthing
site ground
B1505A Interlock
16493J Interlock cable, 3 m or 1.5 m, 1 ea., and 1252-1419
Interlock connector, 1 ea.
Interlock circuit
Shielding box
DUT interface
N1254A-524
Interlock circuit
Sense
Force
Low
High
Sense
Force *2
*1
*2
*3
16493J
N1265A
*1: 1252-1419
*2: N1254A-523
*3: N1254A-506
*4: N1254A-503
*5: 16494A, 6 ea., 16493L, and 16493G
(needs more cables for using HVSMU/HVMCU
or HP/MPSMU through selector of N1265A .)
*5
B1505A
*4
Gate
*4
N1266A
*6
*6: 16494A, 4 ea., 16493L, 16493T, and 16493G to configure HVMCU.
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......