7-76
Keysight B1505A Configuration and Connection Guide
Connection and Ordering Examples
Upgrading from existing B1505A
Example of how to upgrade existing B1505A
•
Existing B1505A system configuration
•
B1505A with one unit of B1512A HCSMU, one unit of B1513A HVSMU, two
units of B1510A HPSMU and one unit of B1520A MFCMU
•
N1259A Test Fixture with N1259A-010 3-line inline socket module option,
N1259A-020 High Voltage Bias-Tee option and N1259A-300 Module Selector
option.
•
Conditions and required specification
•
Device Type: Three terminal discrete MOSFET
•
Max voltage/current: 3 kV/1500 A
•
Capacitance measurement up to 3 kV DC bias
•
Drain leakage current measurement down to 10 fA
Figure 7-39
Connection example of upgraded B1505A
NOTE
The N1259A-010 Inline Package Socket Module (3 pin) cannot be used with the N1265A
Ultra High Current Expander/Fixture due to its current limitation (maximum current is 40
A).
NOTE
The N1259A-020 High Voltage Bias-Tee option cannot be re-used.
Table 7-36
Ordering example
Model/Option
Quantity
Description
B1505AU
1
Power Device Analyzer/Curve Tracer mainframe
B1505AU-014
2
Medium Current Source Monitor Unit (B1514A)
16494A
1
Triaxial Cable
Exis ng B1505A
HVSMU
GNDU
MFCMU
HPSMU
F
S
For Drain
For Drain
For Drain
For Gate
HPSMU
F
S
HCSMU
F
S
New B1505A
For Drain
For Drain
For Gate
F
S
F
S
N1265A
N1265A
Replaced with
2 MCSMUs
Reused
Reused
Reused as a Gate
SMU
Reused
UHC Input
V Control
I Control
N1265A (Selector built-in)
GNDU
HVSMU
Gate F
Gate S
Gate H
Gate L
Force H
Force L
Sense H
Sense L
N1254A-522
(Included in N1265A-010)
N1254A-508
(In clud ed in N12 65A-010)
N1254A-509
(Included in N1265 A-010)
N1265-61751 &
N1265A-612752
(In clude d in N1265 A-010)
Selector
O utput
Selector
Input
16493J-001 Interlock cable (Included in exi ng B1505A)
16493G-001 Digital I/O cable (Included in N1265A)
16493T
16493L-001 (Included in B1505A)
16494A
HVSMU
GNDU
MFCMU
HPSMU
HCSMU
MCSMU
MCSMU
F
S
F
S
SMU H
SMU L
16494A
N1260A
Bias-Tee
F
S
F
S
N1300A
16493S
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......