Keysight B1505A Configuration and Connection Guide
4-15
N1272A and N1273A Connection Guide
Output Connection
Accessories for Connecting a DUT
Use the test lead kit furnished with the N1273A to connect a device that cannot be
connected to any socket module.
See the following simple instruction.
•
Requirements:
•
Test Leads and Connection Kit for Capacitance Test (N1254A-556),
four 20 cm alligator clip - lug cables, four banana plugs, four nuts, and four spare
clips
•
DUT
•
Instruction:
1. Put your DUT on the blank silicon plate.
2. Connect your DUT to the measurement terminals of the test fixture by using the test
leads.
3. Close the fixture cover and perform measurement.
WARNING
Hazardous voltage, instrument maximum output voltage may appear at the
measurement terminals if the fixture cover is closed.
WARNING
Open the AC/DC Guard terminal to avoid electrical shock, instrument damage, or
DUT damage. Do not connect anything to this terminal which may have the same
potential as the Collector/Drain terminal.
WARNING
Make sure that the cover is closed properly before starting measurement. Do not
perform the measurement when a wire is protruding from the fixture cover.