Keysight B1505A Configuration and Connection Guide
7-47
Connection and Ordering Examples
Configuration Examples for Vertical Device Measurement with Wafer Prober
3 kV, 20 A Measurement for On-Wafer Vertical Device
•
Device Type: Vertical MOSFET (3 terminals)
•
Key Specification
•
Max Voltage/Current: 3 kV/20 A
Figure 7-24
Connection example for 3 kV/20A measurement without N1258A Module Selector
Table 7-23
Required equipment, accessories, and cables
Model/Option
Quantity
Description
B1505A
1
Power Device Analyzer/Curve Tracer mainframe
B1512A-FG
1
High Current Source Monitor Unit, 20 A/20 V(Pulsed); 1 A/40 V(DC)
B1513C-FG
1
High Voltage Source Monitor Unit, 3000 V/4 mA (Pulsed & DC)
B1514A-FG
1
Medium Current Source Monitor Unit, 1 A/30 V(Pulsed), 100 mA/30 V(DC)
16493U
1
High Current BNC Coaxial Cable
16493U-001
2
High Current BNC Coaxial Cable (1.5 m)
16494A
1
Triaxial Cable
16494A-003
1
Triaxial Cable (80 cm)
1250-2405
1
BNC-T Plug (m)-BNC (f)-BNC (f) adapter, 1ea
16493S
1
High Current Source Monitor Unit Cable
16493S-011
1
High Current Source Monitor Unit non-Kelvin Adapter
16493S-020
1
Dual HCSMU Combination Adapter
HVSMU
HCSMU
GNDU
MCSMU
B1505A
G
S
HCSMU Non-
Kelvin adapter
16493S-011
Dual HCSMU
Kelvin
Combina
Ɵ
on
adapter
16493S-020
HF
HS
LS
LF
For Gate
N1254A-104 TRX(J) to BNC(P) adapter
F
S
F
S
16493J-001 Interlock cable
D
(Chuck)
For Drain
For Drain
16493T (Included in the HVSMU)
16493S (Included in the HCSMU)
16494A
(Included in the MCSMU)
16493L
(Included in B1505A)
16494A-003
R-Box
1kO
N1262 A-010
1250-2405
16493U-001
R-Box
100kO
N 1262A-002
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......