Keysight B1505A Configuration and Connection Guide
5-35
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect HVSMU Current Expander
Table 5-26
To Connect N1266A Input
CAUTION
The GNDU can sink current of up to 4.2 A. Use 16493L GNDU cable to connect the
GNDU to the N1267A.
Do
not
use normal triaxial cable (16494A) because the maximum current rating of the
cable is 1 A.
Table 5-27
To Connect N1266A Output
CAUTION
Never connect the HVSMU Force and Guard terminals and the HVMCU High terminal to
any output, including circuit common, chassis ground, or any other measurement resource
such as SMU. Connecting other measurement resource may damage the connected one.
B1505A terminals
Required cables
N1266A terminals
HVSMU
Force
16493T-001 HVSMU cable, 1.5 m
HVSMU
GNDU
16493L-001 GNDU cable, 1.5 m
GNDU
MCSMU
1
1. HCSMU is substitutable. Then use 16493S HCSMU cable, N1254A-103 adapter for Force, and
N1254A-517 adapter for Sense.
Force
16494A-001 Triaxial cable, 1.5 m
I Control
Force
Sense
16494A-001 Triaxial cable, 1.5 m
Sense
MCSMU
1
Force
16494A-001 Triaxial cable, 1.5 m
V Control
Force
Sense
16494A-001 Triaxial cable, 1.5 m
Sense
Digital I/O
16493G-001 Digital I/O connection cable, 1.5 m
Digital I/O Input
Power cable, for connecting to power line
LINE
N1266A
terminals
Required cables/Description
Connect to
Low
16493L GNDU cable, N1261A-002 protection adapter and two
BNC(m) coaxial cables (N1254A-503 can be used), or N1261A-004
protection adapter and two SHV(plug) coaxial cables
Protection adapter is not required to connect to the N1258A,
N1259A, N1259A-300, or N1265A.
DUT low terminal
(ex: source)
High
HV(plug) triaxial cable (N1254A-505 can be used).
DUT high terminal
(ex: drain)
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......