3-4
Keysight B1505A Configuration and Connection Guide
N1265A Connection Guide
Input Connection
Input Connection
Prepare the required cables listed in the following tables and connect them between the
B1505A terminals and the relative N1265A terminals.
shows the connections
without HVMCU and UHVU.
shows the additional connections for using
HVMCU. And
shows the additional connections for using UHVU. For the
additional information on the HVMCU, see
"To Connect HVSMU Current Expander" on
. For the additional information on the UHVU, see
Voltage Expander" on page 5-38
Figure 3-1
N1265A Input Connectors on Rear Panel
SMU 1 ([1], [2]), 2 ([3], [4]), and 3 ([5], [6]) connectors are used for connecting the
MPSMU, HPSMU, or MCSMU. Also SMU 3 ([5], [6]) connectors can be used for
connecting HCSMU or DHCSMU. See
for the SMU signal assigned to the SMU
input connector by the type of the connected SMU.
The connectors [1], [3], and [5] are for connecting Force, and the connectors [2], [4], and
[6] are for connecting Sense of the measurement resource connected to the connectors [1],
[3], and [5] respectively.
For connecting HP/MPSMU, the connectors [2], [4], and [6] can be also used to connect
Force of other SMU for non-Kelvin connection.
NOTE
For connecting MFCMU, use N1300A-001 CMU cable, N1260A high voltage bias-tee,
and N1254A-518 SHV(plug)-SHV(plug) cable.
UHV
Bias Tee
SMU
SMU
1
2
3
±10 kV
Max
±3 kV
Max
±200 V
Max 1 A
±200 V
Max
2 A dc 40 A Pulse
To avoid electrical shock and instrument damage,
do not remove the cables during operation.
High
Low
Low
High
To Use HPSMU/MPSMU
Connect the N1261A-001 protection adapter between the SMU
connectors and HPSMU/MPSMU to prevent damage to the SMU.
CAUTION
KCC-REM-ATi-
1H-ParaINSTR
Output
Input
HVSMU
V Control
I Control
MCSMU/HCSMU
MCSMU/HCSMU
SMU
GNDU
Force
Sense
Force
Sense
Force
Sense
Force
Sense
N10149
LINE
100 - 240 V
50/60 Hz
400 VA Max
UHC Input
Interlock
Gate
Digital I/O
Selector Input
±3 kV
Max
±200 V
Max
±40 V
Max
±40 V
Max
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......