4-2
Keysight B1505A Configuration and Connection Guide
N1272A and N1273A Connection Guide
This chapter describes the required information for connecting Keysight B1505A,
N1272A, N1273A, and a device under test (DUT).
•
•
The N1273A capacitance test fixture is a test fixture used for capacitance measurements of
packaged devices. The fixture can be connected to GNDU, MFCMU, HVSMU, and
MPSMU via the N1272A device capacitance selector.
The N1272A selector is connected between the mainframe and the test fixture. And it is
used for switching the IV measurement and the CV measurement, and has a built-in
GNDU protection adapter and a built-in MFCMU protection adapter.
General specifications of the N1272A/N1273A are listed below.
N1272A selector:
•
Dimensions: 420 mm (W)
75 mm (H)
575 mm (D)
•
Weight: 9.4 kg
N1273A fixture:
•
Dimensions: 340 mm (W)
200 mm (H)
345 mm (D)
•
Weight: 4.9 kg
WARNING
To avoid electrical shock and instrument damage, turn the all instruments off before
connecting or disconnecting measurement cable.
WARNING
There are potentially hazardous voltages of up to ± 3000 V (HVSMU) or ± 100 V
(MPSMU) at the Force, Guard, and Sense terminals. To prevent electrical shock, do
not expose these lines.
WARNING
To prevent electrical shock during use, connect the Interlock cable (see
Interlock Circuit" on page 4-3
WARNING
The B1505A is heavy and requires a two person lift.
CAUTION
Do not grab the fixture cover when lifting the N1273A.
CAUTION
Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Connecting anything may damage the module.
CAUTION
Never connect the HVSMU Force and Guard terminals to any output, including circuit
common, chassis ground, or any other module terminal. Connecting other module may
damage the connected module.
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......