3-14
Keysight B1505A Configuration and Connection Guide
N1265A Connection Guide
Output Connection
NOTE
Controlling selector input-to-output path
The configuration and the default I/O path of the selector are set by using the UHC
Expander/Fixture tab screen of the Configuration window of the EasyEXPERT software.
The connection path from the input port to the output port is controlled by each test setup.
And the selector switching status is indicated by the status indicator mounted on the front
panel. See
for the indication and the connection path.
Figure 3-5
Built-in Selector Simplified Internal Connections
Table 3-7
Built-in Selector Status Indicator and Input-to-Output Connection Path
NOTE
Selector may emit a noise sound during operation. However it is not abnormal status.
Measurement resource
used for the test
Status
indicator
Input-to-output connection
1
of output terminals
1. In the normal state, the selector makes the path to the output port from the input port specified by the
Default field. When a measurement is performed, the selector performs automatic switching in every
test. The selector makes the path to the output port from the measurement resource used for the test.
Low Sense
Low Force
High Sense
High Force
2
2. This condition is when the Default field is blank.
(off)
Open
Open
Open
Open
GNDU
GNDU Sense
GNDU Force
Open
Open
HPSMU or MPSMU
SMU
GNDU Sense
GNDU Force
SMU Sense
SMU Force
UHCU
UHC
UHCU
Low Sense
UHCU
Low Force
UHCU
High Sense
UHCU
High Force
HVSMU or HVMCU
HVSMU
GNDU Sense
GNDU Force
HVSMU Force
Open
HVSMU/HVMCU and
built-in series resistor
GNDU Sense
GNDU Force
HVSMU Force
+ Series resistor
Open
Low
High
Sense
Force
S: Sense
F: Force
P.A
GNDU
UHCU
SMU
HVSMU
S
F
S
F
S
F
F
P.A: Protection adapter
Sense
Force
HVSMU force is
connected to
High sense line.
MCSMU MCSMU
I-control
V-control
MCSMU
F
S
Low
High
Gate
100 k
Ω
0
Ω
10
Ω
100
Ω
or 1 k
Ω
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......