7-51
Connection and Ordering Examples
Configuration Examples for Vertical Device Measurement with Wafer Prober
3 kV, 20 A, Capacitance Measurement for On-Wafer Vertical
Device
•
Device Type: Vertical MOSFET (4 terminals)
•
Key Specification
•
Max voltage/current: 3 kV/20 A
•
Capacitance measurement up to 3 kV DC bias
•
Module Selector for multi-parameter automated measurement function
Figure 7-26
Connection example for IV measurement
HVSMU
HCSMU
GNDU
MCSMU
B1505A
16493L (Included in B1505A )
N1258A
D (Chuck)
G
S
HCSMU Non-
Kelvin adapter
16493S-011
HF
LF
For Gate
For Drain
For Drain
16493G-001 Digital I/O cable (Included in N1258A)
N1254A-104 TRX(J) to BNC(P) adapter
F
S
F
S
HVSMU
HCSMU F
GNDU
16493J Interlock cable (Included in B1505A)
N1260A
Bias-T
AC-H
AC-L
A C-Guard
Hc
Hp
Lc
Lp
MFCMU
HCSMU S
N1300A (Included in MFCMU)
16494A (Included in the MCSMU)
16494A-003
16493U-001
1250-2405
16493T (Included in the HVSMU. Please change connec
Ɵ
on manually)
16493S (Included in the HCSMU)
R-Box
1kohm
N1262 A-010
High F
High S
Low S
Low F
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......