7-60
Keysight B1505A Configuration and Connection Guide
Connection and Ordering Examples
Configuration Examples for Vertical Device Measurement with Wafer Prober
3 kV, 500 A, Capacitance Measurement for On-Wafer
Vertical Device
•
Device Type: Vertical Device (3 terminals)
•
Key Specification
•
Max voltage/current: 3 kV/500 A
•
Capacitance measurement up to 3 kV DC bias
•
Module Selector for multi-parameter automated measurement function
Figure 7-32
Connection example for CV measurement
NOTE
Prober vendor is responsible for cabling inside the shielding box.
NOTE
for IV measurement connection example.
NOTE
In this configuration example, a High Voltage Triaxial jack connector is used to connect the
wafer chuck. Then you need a adapter to connect the High Voltage Triaxial plug connector
to the SHV connector. Assemble it by using the N1262A-021 connector or consult your
prober vendor.
16494J Inte rlock cable (Included in B1505A)
16493G-001 Digital I/O cable (Included in N1266A)
G
S
HVSMU
GNDU
MCSMU
B1505A
MCSMU
MCSMU
F
S
F
S
F
S
HVSMU
GNDU
Gate F
Gate S
UHC Input
V Control
I Control
For Gate
For UHC
For UHC
For Drain
N1265A
N1260A
Bias-T
To change connec
Ɵ
on
manually
Hc
Hp
Lc
Lp
HV-F
A C-H
A C-L
AC-Guard
MFCMU
D (Chuck)
N1254A-518
N1300A (Included in the MFCMU)
16494A (Included in the MCSMU)
16493L (Included in B1505A)
16493T (Included in the HVSMU)
N1262A-021
can be used for
custom adapter (SHV jack to
HVTRX jack)
Universal
R-Box
N1262A-021
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......