Keysight B1505A Configuration and Connection Guide
5-37
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect Ultra High Current Expander
Table 5-29
Cables for connecting Prober Side of N1254A-524 System Cable
CAUTION
Never connect the High Force and High Sense terminals to any output, including circuit
common, chassis ground, or any other measurement resource such as SMU. Connecting
other measurement resource may damage the connected one.
Connecting System Cable to N1265A
Requirements
•
Box wrench, M5 hex
•
Clear plastic plate, 2 ea. (furnished with the N1254A-524 system cable)
Procedure
1. Turn the all instruments off.
2. Remove power cable from the all instruments.
3. Open the fixture cover of the N1265A.
4. Remove connection wire from the Selector Output and Gate terminals if the wire is
connected.
5. Remove the clear plastic plate from the fixture cover. This makes opening for passing
the cable through. Use the box wrench and unscrew six hex nuts to remove the plate.
Treasure the removed plate. It must be replaced in the normal use conditions without
the N1254A-524 system cable.
6. Connect two connector boxes of the system cable to the Selector Output and Gate
terminals properly. See
7. Pull out the prober side connector box and some long of extension cable
(approximately 1.5 m) from the opening.
8. Set and fix two clear plastic plates to the fixture cover by using the hex nuts removed in
the step 5. The plates close the excess opening and make the opening enough to pass
the cable through.
9. Close the fixture cover.
10. Connect the 16493J interlock cable from the B1505A or the N1268A to the interlock
circuit installed in the probe station.
WARNING
For using the prober system cable safely
Never connect the interlock cable to the N1265A.
Never remove the clear plastic plates from the fixture cover.
N1254A-524
connectors
Required cables/Description
Connect to
Low Force
Ultra high current cable with banana(plug). N1254A-523 can be used.
DUT low terminal
(ex: source)
Low Sense
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
High Force
Ultra high current cable with banana(plug). N1254A-523 can be used.
DUT high terminal
(ex: drain)
High Sense
Triaxial cable with HV(plug) connector. N1254A-505 can be used.
Gate
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
DUT drive terminal
(ex: gate)
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......