Connection Overview
Table 5-3
To Connect N1258A Module Selector
B1505A terminals
Required cables
N1258A terminals
HPSMU
or
MPSMU
Force
16494A Triaxial cable, 3 m or 1.5 m, 1 ea., for the non-Kelvin
connection, connect it between the Force connectors.
For the Kelvin connection, prepare additional one more cable and
connect it between the Sense connectors. Or use 16493K Kelvin
triaxial cable instead of 16494A.
Use 16493K cable if the High output terminals are connected to
the prober chuck.
HPSMU
Force
Sense
Sense
HCSMU
or
DHCSMU
Force
16493S HCSMU cable, 3 m or 1.5 m, 1 ea. for HCSMU
16493S HCSMU cable, 3 m or 1.5 m, 2 ea. and 16493S-021 Dual
HCSMU adapter, 1 ea. for DHCSMU
HCSMU
Force
Sense
Sense
HVSMU
Force
16493T HVSMU cable, 3 m or 1.5 m, 1 ea.
HVMCU can be used instead of HVSMU. See
HVSMU
GNDU
16493L GNDU cable, 3 m or 1.5 m, 1 ea. Also see
.
GNDU
Digital I/O
16493G Digital I/O connection cable, 3 m or 1.5 m, 1 ea.
Digital I/O
N1258A terminals
Required cables/Description
Connect to
Low
Force
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
DUT low terminal
(ex: source)
1
1. Non-Kelvin connection is not allowed. GNDU Force and Sense are internally connected to Low Sense.
Sense
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
High
Force
Coaxial cable with HV(plug) connector. N1254A-506 can be used.
Use coaxial cable to minimize affect of inductance.
DUT high terminal
(ex: drain)
1
Sense
Triaxial cable with HV(plug) connector. N1254A-505 can be used.
LINE
Power cable, 1 ea., for connecting to power line
site AC power
Shielding box
B 1505A
DUT interface
N1258A
Interlock circuit
16493S
16493T
Sense
Force
Low
High
Sense
Force *2
*1
*4
*1: 1252-1419
*2: N1254A-506 or other cable, 1 ea.
*3: N1254A-505 or other cable, 1 ea.
*4: N1254A-503 or other cable, 2 ea.
16493G
16494A/16493K
16493L
*3
16493J
N1266A
*5
*5: 16494A, 4 ea., 16493L, 16493T, and 16493G
to configure HVMCU.
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......