Keysight B1505A Configuration and Connection Guide
1-39
Configuration Guide
N1271A Thermal Test Enclosure and Related Accessories
Measurement Capability on Thermal Test
The B1505A measurement capabilities supported on the thermal test are summarized in
. Qg tests are not supported on the thermal test.
Table 1-16
Measurement Capability and Requirements
Information of inTEST Thermal Plate and Thermostream
The B1505A with the N1271A supports the Thermal Plate and the Thermostream of
inTEST corporation. Supported models are listed below but contact inTEST for more
information,
.
Capability
Fixture
Thermal Plate
Solution
Thermostream
Solution
•
IV test up to 20 A, 3 kV
N1259A
Yes
(with N1271A-001)
No
•
IV test up to 1500 A, 3 kV
N1265A
Yes
(with N1271A-001)
Yes
(with N1271A-002)
•
IV test up to 1500 A, 10 kV
•
CV test
N1265A
Yes
(with N1271A-001)
Yes
(with N1271A-005)
•
Qg test up to 20 A, 3 kV
N1259A
No
No
•
Qg test up to 1100 A, 3 kV
N1265A
No
No