Keysight B1505A Configuration and Connection Guide
7-49
Connection and Ordering Examples
Configuration Examples for Vertical Device Measurement with Wafer Prober
Figure 7-25
Connection example for 3 kV/20 A automated measurement with N1258A Module
Selector
NOTE
Prober vendor is responsible for cabling inside the shielding box.
Table 7-24
Required equipment, accessories, and cables
Model/Option
Quantity
Description
B1505A
1
Power Device Analyzer/Curve Tracer mainframe
B1512A-FG
1
High Current Source Monitor Unit, 20 A/20 V(Pulsed); 1 A/40 V(DC)
B1513C-FG
1
High Voltage Source Monitor Unit, 3000 V/4 mA (Pulsed & DC)
B1514A-FG
1
Medium Current Source Monitor Unit, 1 A/30 V(Pulsed), 100 mA/30 V(DC)
16493U
1
High Current BNC Coaxial Cable
16493U-001
2
High Current BNC Coaxial Cable (1.5 m)
16494A
1
Triaxial Cable
16494A-003
1
Triaxial Cable (80 cm)
1250-2405
1
BNC-T Plug (m)-BNC (f)-BNC (f) adapter, 1ea
N1258A
1
Module Selector for B1505A
16493S
1
High Current Source Monitor Unit Cable
16493S-011
1
High Current Source Monitor Unit non-Kelvin Adapter
N1262A
1
R-box for B1505A
N1262A-010
1
1 kohm R-box for gate (Triaxial output)
HVSMU
HCSMU
GNDU
MCSMU
B1505A
N1258A
G
S
HCSMU Non-
Kelvin adapter
16493S-011
For Gate
16493G-001 Digital I/O cable (Included in N1258A)
N1254A-104 TRX(J) to BNC(P) adapter
F
S
F
S
High F
High S
Low S
Low F
Digital I/O
HV SMU
HCSMU F
HCSMU S
GNDU
16493J-001 Interlock cable
D
(Chuck)
For Drain
For Drain
16493T (Included in the HVSMU)
16493S (Included in the HCSMU)
16494A (Included in the MCSMU)
16493L (Included in B1505A)
16493U-001
1250-2405
16494A-003
R-Box
1kO
N1262 A-010
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......