![Keysight Technologies B1505A Configuration Manual Download Page 197](http://html1.mh-extra.com/html/keysight-technologies/b1505a/b1505a_configuration-manual_1961941197.webp)
Keysight B1505A Configuration and Connection Guide
5-49
Connection Guide for Wafer Prober and Your Own Test Fixture
To Install an Interlock Circuit
Figure 5-27
Dimensions of the Interlock Switch (Keysight N1254A-402)
To Connect Interlock Circuit
The B1505A provides the Interlock connector to prevent you from receiving an electrical
shock from high voltage (more than ± 42 V). If the interlock circuit is open, the B1505A
cannot
apply high voltage more than ± 42 V.
Before performing measurement, connect the 16493J interlock cable between the B1505A
Interlock connector and the interlock connector which is a part of the interlock circuit
installed in your DUT interface as described in
.
3.1
35.6
8.1
14
NC
NO
C OM
6.4
22.2
27.8
2.8
6.3
15.9 18.8
4.75
10.3
4.3
2.8
2.8
3.2
10.3
2.8
3.1
Switch off
Units: mm
15.3