2-18
Keysight B1505A Configuration and Connection Guide
N1259A Connection Guide
Output Connection
To Use High Voltage Bias-T
The high voltage bias-T is used to perform capacitance or impedance measurements using
DC bias of up to ± 3000 V. The bias-T High terminal is internally connected to the MF
CMU Hpot and Hcur input connectors. And the bias-T Low terminal is internally
connected to the MF CMU Lpot and Lcur input connectors.
•
Required parts:
N1254A-512 SHV(plug)-SHV(plug) cable, 2 ea.
N1254A-513 SHV(jack)-banana adapter, 2 ea.
N1254A-508 or N1254A-509 connection wire, 2 ea.
•
Connection:
See
1. Attach the SHV-banana adapter to the SHV cable, and make two cables.
2. Connect a cable (solid line) between the bias-T Low terminal and the low terminal
of DUT.
3. Connect a cable (solid line) between the bias-T High terminal and the high terminal
of DUT.
4. Connect a wire (dotted line) between a SMU Force output terminal and the DC Bias
Input High terminal. Use MPSMU for DC bias of up to ± 100 V, HPSMU for ± 200
V, or HVSMU for ± 3000 V.
5. Connect a wire (dotted line) between the SMU Guard output terminal and the DC
Bias Input Guard terminal.
Figure 2-9
High Voltage Bias-T Connection Example
High Voltage Bias –Tee
Guard
Module Selector Output
Force
Sense
Guard
Sense
Force
1 MΩ
2
1
100 kΩ
High
Low
2
1
± 3 kV
Max
± 40 V
Max
N1259 A
Opt 022
± 3 kV Max
Max: 6.4 W
± 3 kV Max
Max: 9 W
N1259 A
Opt 021
N1259 A
Opt 010
Force
Sense
Force
Sense
Force
Sense
1
2
3
1 2 3
± 3 kV
Max
Inline Package Socket (3 Pin)
Kelvin Connection
MF CMU
DC Bias Input
± 3 kV
Max
High
Low
Force
± 3 kV
Max
± 25 V
Max
Guard
Guard
Force
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......