5-36
Keysight B1505A Configuration and Connection Guide
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect Ultra High Current Expander
To Connect Ultra High Current Expander
The ultra high current expander is built into the N1265A test fixture, and is used to
configure the ultra high current unit (UHCU) with the MCSMU modules. The UHCU
output can be applied from the Selector Output terminals of the N1265A, and extended by
using the N1254A-524 Prober System Cable, 1.8 m. See
"Selector Output Terminals" on
for the selector itself and the output terminals.
The N1254A-524 system cable is used to extend the Selector Output and Gate terminals of
the N1265A, and can be connected to these terminals directly. See
The other side, prober side, of the system cable is a connector box which provides the five
connectors for connecting cables from manipulators. The connector box should be installed
near your DUT interface. See
for the connection image.
The prober side connector type, maximum voltage, and maximum current of the
N1254A-524 are shown in
for the output connection cables.
CAUTION
Do not grab the fixture cover when lifting the N1265A.
Do not apply voltage/current over the maximum limit of the cable. See
Figure 5-18
To Extend Selector Output by using N1254A-524 Prober System Cable
Table 5-28
Prober Side Connectors and Maximum Voltage/Current of N1254A-524
Connector
label
Connector
type
N1265A output terminals
to extend
Maximum
voltage
Maximum
current
High Force
Banana(jack)
Selector Output High Force
3000 V
39 A dc,
500 A pulse
1
1. Pulse of 1 ms pulse width and 0.4 % duty cycle
Low Force
Banana(jack)
Selector Output Low Force
200 V
High Sense
HV(jack)
Selector Output High Sense
3000 V
1 A dc,
20 A pulse
Low Sense
BNC(f)
Selector Output Low Sense
200 V
Gate Low
Gate
BNC(f)
Gate High
200 V
N1254A-524 Prober System Cable
High Force
Guard
N1265A Test Fixture
Gate
Drain
Source
Selector Output
Chassis
Low Force
High Sense
Low Sense
Gate High
Gate Low
High Force
Low Force
High Sense
Low Sense
Gate
Gate
DUT
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......