AK 98 Dialysis Machine - Attentions
B:181
614 Leakage detected
614 Leakage detected
Internal fluid leakage detected. To continue,
wipe fluid detector container dry.
Appears:
When the machine has detected a fluid leakage.
Appears in:
●
Function check
●
Blood line preparation
●
Pre treatment
●
Post treatment
Machine actions:
●
Close valves to stop leakage to the environment.
Operator action:
1.
Wipe the fluid leakage detector tray dry.
2.
Continue treatment.
Mute time:
2 min