573 Low flow heat is stopped
573 Low flow heat is stopped
To continue to drain phase press Confirm.
Appears:
When the WRO 300 H cannot deliver hot water during
integrated heat disinfection program with WRO 300 H. The
low flow heat phase (where the machine receives hot water
from the WRO 300 H) will not start.
This attention can appear if you manually choose to start a
low flow heat phase but do not confirm it within 5 minutes.
Machine actions:
The low flow heat phase does not start.
Operator action:
1.
Press
Confirm
to confirm the attention. The machine
now enters drain phase.
2.
Contact an authorised technician.
Mute time:
Permanent.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...