HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Alarms
B:43
1.3
Medium priority alarms
200 Arterial pressure limits too wide
200 Arterial pressure limits too wide
Centralize limits.
Appears:
When the arterial pressure alarm window is wider than the
preset value.
Machine actions:
●
Alarm sound is delayed for 2 minutes.
●
The arterial pressure control is activated.
Operator action:
1.
Check that correct blood flow rate is set.
2.
Press the arterial pressure control.
3.
Press
Center arterial
.
4.
Check that the arterial pressure limits are correct.