AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
121 No change in SN venous blood pressure
121 No change in SN venous blood pressure
Check bloodlines and venous pressure
transducer protector.
Appears:
When the venous blood pressure is not varying as expected
during the arterial and venous phases in a single needle
treatment.
Machine actions:
The blood pump stops.
Operator action:
1.
Check the venous blood line for kinks or clamps.
2.
Check for blood in the pressure transducer protector.
3.
Check the arterial blood flow rate.