B:98
AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
523 Dialysis fluid flow is too low
523 Dialysis fluid flow is too low
Appears:
When the flow rate of the dialysis fluid is below the set low
alarm limit. This is because the flow rate is manually set
below the alarm limit or there is an obstruction in the
dialysis fluid flow path.
Machine actions:
None.
Operator action:
If possible, increase the dialysis fluid flow rate. If this is not
possible, contact an authorised technician.
If the attention is not solved, it will appear again in the next
treatment.
Mute time:
Permanent.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...