AK 98 Dialysis Machine - Attentions
B:163
594 Too high TMP required
594 Too high TMP required
Increase time or decrease UFV.
Appears:
When the machine cannot maintain the required UF rate.
Machine actions:
Alarm sound is delayed for 2 minutes.
Operator action:
●
Check if the dialyzer is too small or is clotted, act in
accordance.
●
If possible, adjust treatment time and/or UF volume until
TMP decreases.
Mute time:
2 minutes.