HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions
B:113
539 Fluid tube sensor test
539 Fluid tube sensor test
Attach fluid tubes to safety couplings. To
start disinfection press Confirm.
Appears:
When the dialysis machine cannot detect that the dialysis
fluid tubes are attached to the safety couplings. The
corresponding machine test cannot be made.
Machine actions:
The ongoing disinfection or rinse program is paused.
Operator action:
To continue the disinfection or rinse program:
1.
Check that the dialysis fluid tubes are properly attached
to the safety couplings.
2.
Press
Confirm
to continue the disinfection or rinse
program.
As an alternative, consult an authorised technician.
Mute time:
2 minutes.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...