B:120
AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
546 Functional check is on hold
546 Functional check is on hold
Water temperature is too high.
Appears:
When the inlet water temperature is too high during
functional check.
Machine actions:
The functional check stops. The functional check will
automatically continue when the inlet water temperature
has decreased.
Operator action:
Check possible cause:
●
The inlet water temperature has not decreased below
45 °C within 5 minutes after the high temperature test is
started.
●
The inlet water temperature has not decreased below
39 °C within 5 minutes after the high temperature test is
finished.
If the attention does not disappear, contact an authorised
technician.
Mute time:
Permanent.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...